Appendix A - Technical Specifications

I.

Page 108

1/06/2012 REV 1.10.2

Thermocouple: Per ITS90 Resolution 0.1°C. Thermocouple burnout detection - automatic

Type Range (°C) Accuracy (°C)

Range (°F) Accuracy (°F)

J

-210 to 0°C ± 1.6°C

-340 to 32°F ± 3°F

0 to 1200°C ± 0.7 °C

32 to 2192°F ± 1.2°F

K

-270 to -162°C ± 4.8°C

-454 to -260°F ± 9°F

-162 to -70°C ± 2.2°C

-260 to -94°F ± 4°F

-70 to 1372°C ± 0.7°C

-94 to 2500°F ± 1.2°F

T

-270 to -73°C ± 2.5°C

-454 to -100°F ± 4.5°F

-73 to 200°C ± 1°C

-100 to 400°F ± 1.8°F

E

-270 to -70°C ± 2.5°C

-454 to -94°F ± 4.5°F

-94 to 1000°C ± 0.7°C

-94 to 1832°F ± 1.3°F

N

-270 to 100°C ± 3.0°C

-454 to 212°F ± 5.4°F

100 to 1300 °C ± 0.9°C

212 to 2372°F ± 1.6°F

S

-50 to 0°C ± 2.8°C

-58 to 32°F ± 5°F

0 to 1000°C ± 1.4°C

32 to 1832°F ± 2.5°F

1000 to 1710°C ± 1.1°C

1832 to 3110°F ± 2.0°F

1710 to 1768°C ± 1.4°C

3110 to 3214°F ± 2.5°F

B

0 to 535°C ± 4°C

32 to 1000°F ± 7.2°F

535 to 1800°C ± 2.1°C

1000 to 3300°F ± 3.8°F


U

0 to 100°C ± 0.9°C

32 to 212°F ± 1.6°F

*

R

0 to 260°C ± 3.0°C

32 to 500°F ± 5.4°F

*

260 to 650°C ± 1.4°C

500 to 1200°F ± 2.5°F *

650 to 1760°C ± 1°C

1200 to 3200°F ± 1.9°F *

C

0 to 180°C ± 2.5°C

32 to 360°F ± 2.5°F

*

180 to 1200°C ± 0.8°C

360 to 2220°F ± 1.5°F *

1200 to 2315°C ± 3.6°C

2200 to 4199°F ± 6.5°F *

L

-200 to 0°C ± 1.1°C

-328 to 32°F ± 2°F

*

0 to 890°C ± 0.7°C

0 to 1650°F ± 1.2°F

*

Reference junction compensation accuracy ±0.5 °C (0 °C to 50 °C)

Note: Field calibration improves accuracy to better than +2°F and up to +0.5 and ens compliance with AMS2750D
measurement accuracy requirements for NADCAP accredited suppliers.

RTD: Base accuracy 0.2% or 0.5 °C (1 °F). Resolution 0.1 °C.

2, or 3 wire connection. Cable compensation to +50 Ohm. Open and short-circuit detection.

10 ohm Cu.

-70 to 170 °C

-94 to 338 °F

100 ohm Pt. 385

-220 to 850 °C

-364 to 1560 °F

100 ohm Pt. 392

-180 to 820 °C

-292 to 1500 °F

200 ohm Pt. 385

-220 to 400 °C

-364 to 750 °F

200 ohm Pt. 392

-180 to 400 °C

-292 to 750 °F

100 ohm Ni.

-70 to 300 °C

-76 to 406 °F

120 ohm Ni. 518

-70 to 300 °C

-94 to 570 °F

120 ohm Ni. 672

-70 to 300 °C

-94 to 570 °F

Advertising