5. Lay the IPC assembly on the benchtop with the IPC panel (drain port side) facing up.

• Check the alignment of the glass plates, spacers and clamps. The bottom of the glass

plates, spacers and clamps should be flush. If either glass plate, spacer, or clamp is
not properly aligned or flush, adjust the alignment by loosening the clamps and move
clamps, glass plates and spacers into alignment (Figure 4.3).

• Tighten the clamps by moving the levers back down towards the IPC after the assembly

is flush.

Fig. 4.3. Alignment of glass plate sandwich.

6. To avoid incompatibility problems between combs and spacers after the gel is cast, check the

fit of the combs in the assembled Sequi-Gen GT cell by trying to place them between the plates.

• If the combs clearly will not fit between the plates without damaging the comb, try a

different comb. Optimally, combs should demonstrate slight resistance to being placed
between the glass plates.

4.3 Casting the Gel

Section 7.1 contains a checklist of required items for DNA sequencing. Polyacrylamide

is a hazardous chemical and neurotoxin. Always wear gloves, lab coat, and safety glasses
while working with polyacrylamide.

1. Prepare the gel solution described in Section 7.1 and 7.2.

• Degas the gel solution for 5-15 minutes under a strong vacuum (

26 in./Hg) to insure

reproducible gel porosity.

Table 4.2 Required Gel Volumes Using the Precision Caster Assembly

IPC

0.25 mm

0.4 mm

0.75 mm

0.25 – 0.75 mm 0.40 – 1.2 mm

Size

spacers

spacers

spacers

wedge spacer

wedge spacer

21 x 40 cm

25 ml

35 ml

70 ml

50 ml

60 ml

21 x 50 cm

30 ml

45 ml

90 ml

65 ml

85 ml

38 x 30 cm

40 ml

50 ml

90 ml

38 x 50 cm

55 ml

85 ml

170 ml

120 ml

140 ml

12