Bio-Rad Sequi-Gen GT Sequencing Cell User Manual

Page 18

Advertising
background image

9. When all air bubbles are removed from the tubing, place the luer taper into the injection

port of the precision caster base (Figure 4.6). Tighten the luer taper fitting in place on the
injection port of the precision caster base and begin to slowly inject the gel solution. Slow
and even pressure on the syringe plunger will insure uniform gel casting with no
bubbles (Figure 4.7).

Fig. 4.6. Inserting syringe tubing into precision caster base.

Fig. 4.7. Injecting gel solution into glass plate sandwich.

15