Table 4.3

(continued)

Sequi-Gen GT

Gel

Recommended

Cell Size

Thickness

Power Setting

38 x 30 cm

0.25 mm

70-75 W

38 x 30 cm

0.40 mm

70-75 W

38 x 30 cm

0.75 mm

70-75 W

38 x 30 cm

0.25-0.75 mm

70-75 W

38 x 30 cm

0.40-1.20 mm

70-75 W

38 x 50 cm

0.25 mm

70-80 W

38 x 50 cm

0.40 mm

75-85 W

38 x 50 cm

0.75 mm

75-85 W

38 x 50 cm

0.25-0.75 mm wedge

75-85 W

38 x 50 cm

0.40-1.2 mm wedge

75-85 W

Important: Never allow the gel temperature to exceed 60 °C. Severe damage to the glass
or adhesive bond may result.

Caution: Periodically check the level of the upper buffer to make sure that it is at least
1 cm above the short glass plate.

2. Continue gel electrophoresis until the desired fragment size separation is achieved. Typically,

gel electrophoresis times are monitored by observing the dye front mobility of either the
bromophenol blue (“fast blue”) or xylene cyanol (“slow blue”) during the course of elec-
trophoresis. Fragment and dye front mobility as a function of polyacrylamide percentage are
shown in Table 4.4 below, and should be used as a guide for gel electrophoresis monitoring.

Table 4.4 Migration of Single-stranded DNA in Denaturing Polyacrylamide
Gels in Relation to Dye Marker Gel Migration*

Polyacrylamide Bromophenol

Gel Percentage

Blue

Xylene Cyanol

5%

35 bases

130 bases

6%

26 bases

106 bases

8%

19 bases

75 bases

10%

12 bases

55 bases

* From Ausubel, F. M. et. al., Current Protocols in Molecular Biology, Greene and Wiley, 1993.

4.7 Disassembly

1. When the desired dye front mobility has been achieved, turn off the power supply, and

remove both safety covers.

• The upper buffer chamber can be partially emptied by inserting the drain port

connector (and any attached tubing) into the drain port on the IPC. A “click” will be
heard when the drain port/tubing connector has been properly inserted (Figure 4.10).

• Buffer will begin to drain from the IPC immediately after the connector is inserted

into the drain port.