Total signal artifacts

Many artifacts appear in more than one sub-category. For example, smile patterns can

exist in lane-local or in total signal situations (or both), but rarely appear in set-local. It is
important to distinguish the extent and location of each artifact before trying to analyze or
troubleshoot the anomalous pattern.

1. Lane-local artifacts

Condition Probable

Causes

Solutions/Preventions

2. Set-local artifacts

Condition Probable

Causes

Solutions/Preventions

Smiling within sets

Frowning within sets

Funneling within sets

Non-continuous vertical
register

Band spacing compressed within
a set

Film exposure differences within
a set

• Loading wells not straight

within set

• Molarity problems in the sam-

ples of the set

• Loading wells not straight
• Molarity problems in the sam-

ples of the set

• Sample molarity too high, or

contamination

• Sample dependent artifact
• Autoradiographic artifact

• Sample dependent artifact

• Sample dependent artifact

• Pour new gel with better lanes

• Check sample and gel buffers

• Pour new gel with better lanes
• Check sample and gel buffers

• Check sample and gel buffers

Smiling bands

Frowning bands

Complex (curvy) bands

Funneling or lane narrowing

Lane widening

Variations in signal intensity
along the lane(s)

Band spacing compressed or
stretched

• Loading wells not straight

• Sample molarity too high

relative to gel

• Sample overloaded
• Sample was allowed to diffuse

into gel too long before
electrophoresis

• Loading wells not straight

• Sample molarity too low rela-

tive to gel

• Sharkstooth comb inserted too

far into the gel

• Loading wells not straight or

clean

• Bubble in gel

• Sample molarity too high

• Sample molarity too low

• Sample dependent artifact
• Check sample and gel buffers

• Sample dependent artifact

• Allow gel to polymerize more

completely before removing
comb

• Check sample and gel buffers

• Reduce sample load volume
• Load a set or two at a time

• Pour new gel with better lanes

• Check sample and gel buffers

• Touch the top of the gel with

the teeth

• Allow gel to polymerize more

completely before removing
comb

• Pour new gel without bubbles

• Check sample and gel buffers

• Check sample and gel buffers