Bio-Rad Model 550, Ultramark, and Benchmark Microplate Reader Accessories User Manual

Page 17

Advertising
background image

4.3 Detailed Operation

Set Analysis Mode

In this series of screens the user chooses single or dual wavelength measurement, assigns

the measurement and reference (for dual wavelength measurement) filters, and selects or des-
elects mixing.

1. Enter the SET ANALYSIS mode by pressing MODE until the display reads

2. Press either ARROW until the display reads

3. Press ENTER. The display will read

4. Dual/Single Wavelength Setting:

a. Use either ARROW to move the cursor under the check box next to D/S.

b. Press SELECT to darken the check box next to D/S.

c. Press ENTER. The display changes to read

d. Use either ARROW to move the cursor under the check box next to the desired

mode.

e. Press SELECT to darken the check box next to the choice.

f. Press ENTER to finalize the selection and the display will return to the previous

screen.

14

>PRINT REPORT

MODE

>SET ANALYSIS

MODE

Set Analysis

D/S,

Filt and

Mix

Select D/S:

Dual/

Single

Set Analysis

D/S,

Filt and

Mix