7.2 RS-232 Interface Specifications

The following specification describes the syntax and language required for computer con-

trol of the reader. In addition, the Model 550 reader will perform passive unidirectional data
transmission. The instrument will automatically transmit the data in ASCII code after each
reading if it is properly connected to a host computer. See the “RTPLATE” command for an
example of ASCII output.

Note: This automatic output will be of Measurement O.D. minus Reference O.D. when

in Dual Wavelength mode (i.e. only one set of 96 numbers).

1. SIGNALS

TXD: transmit data
RXD: receive data
DTR: data terminal ready
DSR: data set ready
CTS: clear to send
RTS: ready to send
SIGNAL GROUND (SHIELD GROUND to hood / chassis ground)

2. TRANSMISSION

Baud rate

9600

Data width

8

Stop width

1

Parity None

3. DATA

All data must be transmitted in ASCII.
The READER command interpreter must be “case blind” (no distinction between
upper or lower case ASCII characters).
All READER commands can be distinguished from each other by the first two
characters of the command.

4. SIGNAL ASSIGNMENT

Pin #

READER Signal type

Host Signal Pair

1

SHIELD GROUND

SHIELD GROUND

2

TRANSMIT DATA

RECEIVE DATA

3

RECEIVE DATA

TRANSMIT DATA

4

REQUEST TO SEND

CLEAR TO SEND

5

CLEAR TO SEND

REQUEST TO SEND

6

DATA SET READY

DATA TERMINAL READY

7

SIGNAL GROUND

SIGNAL GROUND

20

DATA TERMINAL READY

DATA SET READY

Shield ground

1

Transmit data

2

Receive data

3

Data set ready

6

Signal GND

7

Data terminal ready

20

Clear to send

5

Request to send

4

32

Host

Computer

Model 550

Reader