System, Legal information, Android version – Bio-Rad ZOE™ Fluorescent Cell Imager User Manual

Page 14: Kernel version, Vision board drive version, Driver board firmware version, Galileo software version, Firmware update, System test posting

Advertising
background image

ZOE Fluorescent Cell Imager Instruction Manual

6 |

System

The following information is available in this menu option:

n

Legal information

n

Android version

n

Kernel version

n

Vision board drive version

n

Driver board firmware version

n

Galileo software version

n

Firmware update

n

System test posting

Firmware update

To update firmware on the instrument:
1. Insert a USB flash drive with the firmware update file into USB port. The firmware file must

be saved in the root directory of the USB flash drive

2. Tap Firmware update in the System option of the Settings menu.
3. When the instrument recognizes and validates the firmware file, it proceeds with the update.
Note: If the USB key with firmware file is not inserted, a message prompting you to do so
will be displayed. If the update file is not saved in the root directory, a “No update file present.”
message displays.
4. Tap Enter to continue updating.
5. When the update is successfully completed, the ZOE Cell Imager automatically restarts.
Note: Three types of update files may be available: a Vision board driver update, Driver board
firmware, and/or ZOE software.

If both the Vision board driver and ZOE software are updated at the same time, the Vision board
driver will be updated first. When that is completed the instrument automatically restarts itself
and repeats steps 2–5 to update the ZOE software.

If the Driver board firmware is also present, it and any other update files are automatically
updated without repeating steps 2–5.