Bio-rad laboratories resources – Bio-Rad ZOE™ Fluorescent Cell Imager User Manual

Page 4

Advertising
background image

ZOE Fluorescent Cell Imager Instruction Manual

ii |

ii |

Bio-Rad Laboratories Resources

Bio-Rad provides many resources for scientists. Bio-Rad resources and how to locate what you need are listed below.

Bio-Rad resources.
Resource

How to Contact

Local Bio-Rad Laboratories

Find local information and contacts on the Bio-Rad Laboratories website by selecting your country

representatives

on the homepage (www.bio-rad.com). Find the nearest international office listed on the back of
this manual

Technical support scientists

Bio-Rad’s technical support scientists provide our customer with practical and expert solutions.
To find local technical support on the phone, contact your nearest Bio-Rad office. For technical
support in the United States and Canada, call 1-800-424-6723 (toll-free), and select the technical
support option

Service support engineers

Maintenance and repairs should be carried out only by authorized service support engineers

For service support in the United States and Canada, call 1-800-424-6723 (toll-free), and select
the technical support option to request service support

Technical notes and literature

Go to the Bio-Rad website (www.bio-rad.com). Type a term in the Search box and select
Documents tab to find links to literature

Warranty

The ZOE Fluorescent Cell Imager and associated accessories are covered by a standard Bio-Rad warranty. Contact your local Bio-Rad
Laboratories office for the details of the warranty.

Safety Use Specifications and Compliance

A

Warning! label in this manual warns you about sources of injury or harm, including risk of electrical shock.

Warning! Do not attempt to repair or remove the outer case of the ZOE inverted cell imager or other accessories. If you open this
instrument, you put yourself at risk for harm to body or equipment from electrical shock.

This instrument has been tested and found to be in compliance with all applicable requirements of the following safety and
electromagnetic standards.

Environmental conditions for safe operation of the ZOE

Fluorescent Cell Imager.
Transient category

Installation category II

Operating power

100–240 V AC

Frequency

50/60 Hz

Electrical input

24 V DC, 100 W

Installation site

Indoor use only

Operating temperature

10–31°C

Maximum relative humidity

20–80%

Altitude

<2,000 m

Pollution degree

2

Degree of protection

IPX0

Note: Do not store or operate the unit near a sink as contact with water could cause electric shock.

Unit is heavy, do not store or operate it at the edge of a laboratory bench. Unit must be in upright position during operation.

Condensation will form on any object when the temperature of the object is at or below the dew point temperature of the air
surrounding the object. If the instrument has been taken from a cold environment to a warmer environment, allow the instrument to
equilibrate to above the dew point temperature before operating.

Do not tip the unit over while a sample is located on the sample stage. This could cause accidental spillage of sample into the ZOE
Fluorescent Cell Imager’s optical and electronic systems, causing electrical shock and rendering unit inoperable.

Do not operate if the glass layer on the LCD screen is broken; this could injure the operator.

To prevent injury from movement of motorized stage, do not put fingers into the stage opening while the unit is powered on.

The angle on LCD screen is adjustable, keep fingers away from the back of the screen when folding down the screen. The ZOE
Fluorescent Cell Imager is factory calibrated; no further calibration is needed.

Do not connect the ZOE Fluorescent Cell Imager to a PC, it may render the device’s operating system unusable.

Advertising