Optimizing Imaging Parameters for Fluorescence Imaging

Use the Edit menu to adjust these precapture parameters (Figure 7): Gain, LED Intensity, and
Contrast values; Exposure time (msec); and quadrant illumination. Drag the slider bar to adjust a
parameter. Tap the – and + controls at either end of each slider bar for fine adjustments.

n

When searching for sample — increase the Gain value for brighter signal; reduce the
Exposure time to obtain a faster frame rate

n

Once the sample is identified and ready to be captured — lower the Gain value
to reduce the background noise; increase the Exposure time to regain signal intensity
as needed

n

To further increase brightness — start by increasing LED intensity value; follow by
increasing Exposure time if needed

Fig. 8. Samples pseudo-colored in the appropriate emission color in Live mode. A, BPAE cells, mitochondria stained with

MitoTracker red CMXRos; B, BPAE cells, F-actin stained with Alexa Fluor 488 phalloidin.

A B

Fig. 7. Edit controls in fluorescent channels.

Gain: 1

Apply

Cancel

Exposure: 50

LED Intensity: 50

Contrast: 50