ZOE Fluorescent Cell Imager Instruction Manual

| 9

Focus position indicator on the left side of the touch screen shows the Z axis position number.
The number is updated during focusing (Figure 4). Knowing the approximate Z number for the
tissue culture vessels (Table 2) you use can speed up focusing.

Table 2. Z axis position range of tissue culture vessels.
Vessel Type

Z Axis Position

Microscope slide

720 (± 100)

6-well plate

2,000 (± 100)

12-well plate

2,100 (± 100)

24-well plate

1,900 (± 100)

48-well plate

2,000 (± 100)

96-well plate

3,500 (± 200)

T25 flask

1,800 (± 200)

T75 flask

2,500 (± 200)

T225 tissue culture flask

3,200 (± 200)

Small petri dish

1,200 (± 200)

Large petri dish

1,100 (± 200)

Optimizing Imaging Parameters in Brightfield Channel

Use the Edit menu to adjust these precapture parameters (Figure 5): Gain, LED Intensity, and
Contrast values; Exposure time (msec); and quadrant illumination. Drag the slider bar to adjust a
parameter. Tap the – and + controls at either end of each slider bar for fine adjustments.

n

When searching for sample — increase Gain value for brighter signal; reduce Exposure
time for faster frame rate

n

Once the sample is identified and ready to be captured — lower Gain value to reduce
background noise; increase Exposure time to regain signal intensity

n

To further increase brightness — start by increasing LED Intensity value; follow by
increasing Exposure time if needed

n

To reduce nonspecific background signal — increase Contrast value

Fig. 5. Edit controls in the brightfield channel.