Fig. 4.3 Rear Panel connections.
4.3 Rear Panel Functions
Connection
Function
Power entry
Grounded 3-pin receptacle for the
module
power cord. Also contained within is
a four-position 100 V/120 V/ 220 V/
240 V line voltage selector. The
collector is shipped in its 120 V or
220 V version. For operation at other
voltages, refer to Section 3.2.
9-pin “D”
For remote operation of the collector.
connector
The Model 2110 can be controlled by
Bio-Rad’s Model EP-1 Econo Pump,
the BioLogic LP low pressure
chromatography system, or the
BioLogic System, a high resolution
liquid chromatography system. The
Model 2110 can also be controlled by
other non-Bio-Rad equipment. See
Section 7 for additional information.