Section 9
Cleaning and Maintenance

The Model 2110 Fraction Collector requires very little maintenance to

insure reliable operation. Two procedures are outlined below.

9.1 Drop Sensor

The Model 2110 Fraction Collector counts drops through an optical

sensor located in the drop area inside the drop former arm. Precipitates or
other foreign debris may obscure the optical path and cause erroneous
drop counts. To prevent this problem, the drop sensor should be cleaned
periodically. An ordinary cotton swab moistened with water can be used
to clean the sensor.

9.2 Instrument Case, Carousel, and Drainage Tube

During normal operation, spills and splashes may result in precipi-

tates forming on the case and carousel. Use a squirt bottle and soapy water
to wash down carousel and case. The water will drain out the back of the
unit through the drainage tube. Prior to squirting the instrument case,
make sure the rubber seal around the gear device is in place. Insure that
the waste tube is draining properly to avoid instrument damage.

Section 10
Troubleshooting

Problem

Cause

Remedy

Drops miss tubes

Carousel misaligned

Reposition carousel

Drop former missing

Replace drop former

Drop head misaligned

Align

Unit not level

Place unit on level surface

Drops miss tube

Flow rate greater

Reduce flow rate

during advance

than 5 ml/min

Drops not counted Drop sensor is dirty

Clean drop sensor

or E08 error code

Flow rate greater than Use Time collection mode
5 ml/min
Operating in a windy

Try relocating unit, or put

environment

a piece of tape over slot in
drop arm

No power

Fuses

Check fuses, replace if
necessary

19