Error
Number
Cause
Remedy
E00
Hardware failure
Unplug and replug to
E01
mains power. If error per-
E04
sists, contact your local
Bio-Rad representative.
E02
Motor jam
Reposition carousel;
unplug and replug to
mains power. If problem
persists, apply WD-40 to
the motor shaft. If motor
must be replaced, order
part number 8006484.
E08
Drop sensor
If problem persists after
E10
cleaning the drop sensor
with a wetted cotton
swab, follow the
Procedure for
Adjusting Drop Detector
Signal Level. See
Appendix B. If drop sen-
sor must be replaced,
order part number
8006481.
Appendix A
Technical Specifications
Capacity
Eighty 13 x 100 mm glass or plastic test
tubes, or 1.5 ml micro test tubes, when
using adaptor (catalog number 731-8135)
Collection basis
1 to 999 drops in 1 drop increments
0.05 to 9.99 minutes in 0.01 minute
increments
10.0 to 99.9 minutes in 0.1 minute
increments
See Econo Pump manual for other collec-
tion options when connected to the Econo
System
Materials of construction
Case and carousel: polypropylene
Front panel: hard-coated polycarbonate
Drop former: silicone rubber
Waste outlet tube: polyurethane