Slurry packing an imac column, Recommended columns, Materials – Bio-Rad Nuvia™ IMAC Resin User Manual

Page 15

Advertising
background image

Nuvia IMAC Ni-Charged Resin 11

Section 4
Column Packing — Medium-Pressure Columns

Slurry Packing an IMAC Column

Slurry packing is preferred for small columns. This method
describes packing a Nuvia

IMAC resin with aid from a pump. For

best results, use 5–50 mm ID columns and a bed height of
5–30 cm.

Recommended Columns

Bio-Rad’s Bio-Scale

MT high-resolution columns may be used

for the following column-packing procedure. These columns are
empty and may be packed with the support of choice.
Bio-Scale MT columns are convenient to use with Bio-Rad’s
NGC

chromatography system or any medium- or high-pressure

system:
• Bio-Scale MT2 column (7 x 52 mm) for bed volumes

up to 2 ml

• Bio-Scale MT5 column (10 x 64 mm) for bed volumes

up to 5 ml

• Bio-Scale MT10 column (12 x 88 mm) for bed volumes

up to 10 ml

• Bio-Scale MT20 column (15 x 113 mm) for bed volumes

up to 20 ml

Materials

• Empty column (1–5 cm ID x 30 cm) with flow adaptors, inlet

and outlet ports

• Glass filter
• Nuvia IMAC Ni-charged resin
• Packing reservoir
• Pump

Advertising
This manual is related to the following products: