Bio-Rad Profinity IMAC Resins User Manual

Page 21

Advertising
background image

Table 6. Purification Method Suggestions

1 ml Cartridge

5 ml Cartridge

Step

CV

Flow Rate

Flow Rate

Equilibrate

5

2 ml/min

10 ml/min

Lysate Load

5 to 10

2 ml/min*

10 ml/min*

Wash 1

6

2 ml/min

10 ml/min

Wash 2

6

2 ml/min

10 ml/min

Elute

5

2 ml/min

10 ml/min

Standard methods that are compatible with any
type of chromatography system are listed below. To
maximize binding capacity with large proteins
(>100 kD), for purification at 4°C, or for purifications
under denaturing conditions, the lysate load flow
rate* can be decreased (to 0.5 ml/min for the 1 ml
cartridge and 2 ml/min for the 5 ml). This will have
to be determined empirically for individual proteins.

1.

Equilibrate the cartridge with 5 column volumes
(CV) of equilibration/wash buffer 1 at 2 ml/min.

2.

Load the sample lysate at 2 ml/min.

3.

Wash the cartridge with 6 CV of wash buffer
1 at 2 ml/min.

18

10005455A.qxp 3/27/2006 2:42 PM Page 18

Advertising
This manual is related to the following products: