Section 11
Troubleshooting Guide

Problem

Possible Cause

Solution

Cartridge clogging

Particulates in samples

Filter all samples and buffers

or slow flow rate

or buffers

through 0.2 µM filter prior to
application

Sample too viscous

Add nuclease to lysate to
degrade DNA. Centrifuge
and filter lysate again

No target protein

Low level of target

Check expression level by

in eluant

protein in starting

SDS-PAGE

material

Target protein not

Check levels of target protein

binding, or eluting in

in lysate, flow-through, wash,

wash fractions

and eluted fractions. Check
for presence of His-tag with
anti-His antibody

Target protein in

His-tag not accessible

Reclone His-tag onto opposite

flow through

terminus (N or C-terminus)

His-tag not accessible

Purify protein under denatur-
ing conditions to expose His-
tag

Proteolysis & removal

Include protease inhibitors in

of tag

lysis buffer (or reaction), or
purify in the cold

25

10005455A.qxp 3/27/2006 2:42 PM Page 25