background image

The cartridges can be used with any liquid
chromatography system capable of setting a high
pressure limit of 45 psi (equivalent to 3 bar or 300 kPa).
Alternatively, luer fittings offer convenient connection
directly to a Luer-Lok syringe for quick, one-step
purification.

Table 1 Bio-Scale Mini IMAC Cartridge
Specifications

Sizes

1 ml and 5 ml bed volumes

Dimensions

1 ml: 40 mm length x 5.6 mm inner
diameter

5 ml: 40 mm length x 12.6 mm
inner diameter

Maximum pressure tolerance

45 psi

Recommended flow rates

1 ml: 1–2 ml/min (240 cm/hr–480 cm/hr)

5 ml:5–10 ml/min (240 cm/hr–480 cm/hr)

Maximum flow rate

1 ml: 6 ml/min (1440 cm/hr)
5 ml: 20 ml/min (963 cm/hr)

Fittings

Female luer-lock inlet and
male luer-lock outlet

Column material

Polypropylene

Frit material

Polyethylene (HDPE)

Shipping conditions

20% ethanol

Storage recommendations

20% ethanol

Autoclavability

Not autoclavable

3

10005455A.qxp 3/27/2006 2:42 PM Page 3

Advertising