Bio-Rad Profinity IMAC Resins User Manual

Page 7

Advertising
background image

Table 2. Profinity IMAC Resin Specifications

Functional ligand

IDA

Base bead

UNOsphere

Particle size range

45–90 µm

Mean particle size

60 µm

Metal ion capacity

12–30 µmol Cu

2+

/ml

Dynamic binding capacity*

≥15 mg/ml

Recommended linear flow rate

480 cm/hr

Maximum operating pressure

45 psi

Chemical compatibility

See table

Storage

4°C to ambient temperature

Shelf life in 20% EtOH

>1 year at ambient temperature

Operational temperature

4–40°C

*

Q10% determination of 1.8 mg/ml (His)6-tagged pure protein (32 kD).

Note: Dynamic binding capacity will vary from protein to protein.

Profinity IMAC cartridges are compatible with aqueous
buffers most commonly used with IMAC purification
techniques. For a complete list of chemical
compatibilities, refer to the Profinity IMAC
instruction manual, online at http://www.bio-
rad.com/cmc_upload/Literature/164111/10001677
B.PDF.

4

10005455A.qxp 3/27/2006 2:42 PM Page 4

Advertising