Bio-Rad Profinity IMAC Resins User Manual
Page 26
Advertising
![background image](/manuals/600412/26/background.png)
For the 1 ml cartridges, run the cleaning protocol at
2 ml/min. It is recommended that the 5 ml cartridge
cleaning protocol be run at 5 ml/min.
High Salt/Acid Cleaning
1.
Rinse the cartridge with 2 CV water at 2 ml/min.
2.
Wash the cartridge with 5 CV 500 mM NaCl,
50 mM Tris, pH 8.0 at 2 ml/min.
3.
Wash the cartridge with 5 CV 500 mM NaCl,
100 mM NaOAc, pH 4.5 at 2 ml/min.
4.
Rinse the cartridge with 2 CV water at 2 ml/min.
5.
Store the cartridge in 20% ETOH at 4°C.
Chaotrope Cleaning
1.
Rinse the cartridge with 2 CV water at 2 ml/min.
2.
Wash the cartridge with 5 CV 6 M guanidine
HCl at 2 ml/min.
3.
Rinse the cartridge with 2 CV water at 2 ml/min.
4.
Store the cartridge in 20% ETOH at 4°C.