For the 1 ml cartridges, run the cleaning protocol at
2 ml/min. It is recommended that the 5 ml cartridge
cleaning protocol be run at 5 ml/min.

High Salt/Acid Cleaning

1.

Rinse the cartridge with 2 CV water at 2 ml/min.

2.

Wash the cartridge with 5 CV 500 mM NaCl,
50 mM Tris, pH 8.0 at 2 ml/min.

3.

Wash the cartridge with 5 CV 500 mM NaCl,
100 mM NaOAc, pH 4.5 at 2 ml/min.

4.

Rinse the cartridge with 2 CV water at 2 ml/min.

5.

Store the cartridge in 20% ETOH at 4°C.

Chaotrope Cleaning

1.

Rinse the cartridge with 2 CV water at 2 ml/min.

2.

Wash the cartridge with 5 CV 6 M guanidine
HCl at 2 ml/min.

3.

Rinse the cartridge with 2 CV water at 2 ml/min.

4.

Store the cartridge in 20% ETOH at 4°C.