Bio-Rad Bio-Scale™ Mini Macro-Prep® High Q and Macro-Prep High S Cartridges User Manual

Page 18

Advertising
background image

4. Equilibrate the cartridge with low-salt buffer for

5 min.

5. Reduce the flow rate to that which will be used

in the purification protocol.

4.1 Sample Preparation

Proper pH and ionic strength are necessary for
consistent and reproducible results. Sample can be
exchanged into the starting buffer or diluted to the
starting buffer's concentration. This can be
achieved by diluting the sample to the ionic strength
of the starting buffer, dialyzing against the starting
buffer, or exchanging it into the starting buffer.
Buffer exchange can be accomplished using a
Bio-Scale Mini P-6 cartridge, Bio-Spin

®

6 or

Bio-Spin 30 column, Econo-Pac

®

10DG desalting

column, or Bio-Gel

®

P-6DG gel filtration gel (see

Table 3). The choice of product will depend on
sample volume. All samples should be filtered
through a 0.45 µm filter prior to cartridge application.

14

10006400B.qxp 7/10/2006 3:48 PM Page 14