Section 2
Connecting to Bio-Rad’s
Low-Pressure
Chromatography
Instruments
Bio-Scale Mini cartridges are ideal for use with
Bio-Rad's BioLogic
™
LP chromatography system,
Econo
™
gradient pump, the patented* Model EP-1
Econo
™
pump, and all low-pressure chromatogra-
phy instruments. Bio-Scale Mini cartridges can be
conveniently connected directly to the system using
the luer fittings on the cartridge.
1. Install 1.6 mm inner diameter (ID) tubing in the
pumphead. Adjust the platen pressure screw
(on pumphead) – using a screwdriver or coin,
turn the screw counterclockwise as far as it will
go, then turn clockwise three full turns.
* US patent 5,135,658