9. Plot all 8 spectra. Using the peak selection tool, select the insulin peak

(5.963 kD) in all spectra.

10. Export the data according to the instructions in Appendix A and

selecting the parameters for export in the following order:

Array bar code

Spot name

Substance mass

Intensity

Resolution

S/N

11. Paste the data into the ProteinChip SELDI system check form, data

sheet “Test 2 Mass Drift and Resolution”.

12. The spreadsheet indicates whether the test passed or failed.

13. If the test fails, run the HV conditioning procedure (maintenance

procedure 1) and repeat the test. Contact technical support if the
test fails a second time.