![background image](/manuals/600506/6/background.png)
Table 1. Summary of procedures and tests (
continued).
C
Co
on
nssu
um
maab
bllee
E
Essttiim
maatteed
d T
Tiim
mee
T
Teesstt
U
Usseed
d
R
Reeq
qu
uiirreed
d
D
Deessccrriip
pttiio
on
n
T
Teesstt 2
2
Mass drift and
ProteinChip
20 min to 1 hr
This test uses the
resolution at
peptide
ProteinChip peptide
5.96 kD
standard array
standard array to measure
the mass drift and
resolution of insulin. These
measurements are
compared to a specification,
and a pass/fail disposition
is obtained.
1.2 Storage and Handling
Store the arrays supplied in the kit in the original packaging at room
temperature and in a dry and dark location. Each array is individually
wrapped. Do not open the array packaging until ready to use. Take care
when handling the ProteinChip arrays to avoid contaminating or disturbing
the matrix crystals on the surface of the arrays.