DFI INFINITY CS (V- M3598/N/0906)

17

CF 70

You may use Reading Configuration

to configure your meter for the following:

• To select ratiometric or non-ratiometric operation

• To set the input resolution of your meter

• To display the filtered/unfiltered signal input value

• To select gross/net vs. peak reading

4.4.1 Selecting Ratiometric/Non-Ratiometric Operation

1. Press

the

MENU button until

displays.

2. Press

the

X/TARE button. The meter flashes one of the following:

(Ratiometric reading) (Default – for strain meters)

(Non-ratiometric reading – typically for voltage & current transducers)

3. Press

the

S/NT/GRS button to view the last stored selection and to toggle between selections.

4. Press

the

X/TARE button to select input resolution or press the MENU button to store your selection and

shows

menu.

4.4.2 Setting Input Resolution

To set the input resolution of your meter, follow these steps:
1. Press

the

MENU button until

displays, then press the

X/TARE button twice.

or
Press the

X/TARE button from

.

One of the following displays (default is

):

= 10µV Unipolar inputs. 25 µV for bipolar inputs.

= 1µv for Unipolar inputs. 5 µV for bipolar inputs.

= 2µV for Unipolar inputs. 10 µV for bipolar inputs.
= 3µV for Unipolar inputs. 15 µV for bipolar inputs.
= 5µV for Unipolar inputs. 25µV for bipolar inputs.


Example: 3 µV resolution means that if you input 0-30 mV, at 30 mV the meter shows

.


2. Press

the

S/NT/GRS button to scroll through available selections.

3. Press

the

X/TARE button to display the filtered/unfiltered signal input or press the MENU button to store your

selection and shows

menu.

4.4.3 Displaying the Filtered/Unfiltered Input Signal

To display the filtered/unfiltered signal input, follow these steps:
1. Press

the

MENU button until

displays, then press the

X/TARE button three times.

or
Press the

X/TARE button from

.

One of the following displays:

= (Filtered value) (Default)

= Unfiltered value

2. Press

the

S/NT/GRS button to toggle between available choices.

3. Press

the

MENU button to store your selections.

momentarily displays, followed by

Menu.

4.4.4 Selecting Gross/Net or Peak Display

To select gross/net or peak display.
1. Press

the

MENU button until

displays, then press the

X/TARE button three times.

or
Press the

X/TARE button from

.

One of the following displays:

= Gross/Net Display (Default)

= Peak Display

2. Press

the

S/NT/GRS button to toggle between available choices.

3. Press

the

MENU button to store your selections.

momentarily displays, followed by

Menu.