Cooper Instruments & Systems DFI Infinity CS Economy Digital Indicator User Manual

Page 22

Advertising
background image

DFI INFINITY CS (V- M3598/N/0906)

19

CF 70

= Setpoint 1 assigned to net reading (Default)

= Setpoint 1 assigned to gross reading

2. Press

the

S/NT/GRS button to toggle between available choices.

3. Press

the

MENU button to store your selection(s). The meter momentarily shows

, followed by

(Setpoint 2 Configuration).

4.7 Using Setpoint 2 Configuration

Refer to Table 6-1 for a summary list of menu configuration.

Setpoint 2 Configuration

is not active unless your meter has dual relay output capabilities. The LED’s will

display whether the

is active or not. You may use Setpoint 2 Configuration

for the following:

• To set setpoint’s active band above or below your chosen value

• To select whether the setpoint operation is latched or unlatched

• To assign setpoint values to the net or gross reading

4.7.1 Setting Setpoints 2’s Active Band

1. Press

the

MENU button until the meter shows

.

or

2. Press

the

X/TARE button. The meter shows one of the following:

= (Active above the setpoint) (Default)

= (Active below the setpoint)

3. Press the

S/NT/GRS button to toggle between available choices.

4. Press the

X/TARE button to select if Setpoint 2 is latched/unlatched or press the MENU button to store your

selection. The meter shows

.

4.7.2 Selecting if Setpoint 2 is Latched or Unlatched

1. Press

the

MENU button until

displays, then press the

X/TARE button twice.

or
Press the

X/TARE button from

.

The meter shows one of the following:

= Setpoint 2 to be unlatched (Default)

= Setpoint 2 to be latched

2. Press

the

S/NT/GRS button to toggle between available choices.

3. Press

the

MENU button to store your selection(s). The meter momentarily shows

, followed by

(Setpoint 1 Deadband).

4.7.3 Assigning Setpoint 2 Values to Net or Gross Readings

1. Press the MENU button until

displays, then press the

X/TARE button twice.

or
Press the

X/TARE button from

.

The meter will show one of the following:

= Setpoint 2 assigned to net reading (Default)

= Setpoint 2 assigned to gross reading

2. Press the

S/NT/GRS button to toggle between available choices.

3. Press the MENU button to store your selection(s). The meter momentarily shows

, followed by

(Setpoint 1 Deadband).

4.8 Setting the Setpoint 1 Deadband

Refer to Table 6-1 for a summary list of menu configuration.

Setpoint 1 Deadband

is not active unless your meter has dual relay output capabilities. The LED’s will

display whether the

is active or not. The Setpoint 1 Default deadband is 0003. To change the deadband

(hysteresis) of Setpoint 1, follow these steps:

Advertising