Cooper Instruments & Systems DFI Infinity CS Economy Digital Indicator User Manual

Page 25

Advertising
background image

DFI INFINITY CS (V- M3598/N/0906)

22

CF 70

equal to the meter’s display and/or any engineering units you require. You may scale the output for direct (4-20
mA, 0-10 V, etc) or reverse acting (20-4 mA, 10-0 V, etc).

1. Press

the

MENU button until

displays.

2. Press

the

X/TARE button.

(Read 1) displays.

This is your first point of display reading.

3. Press

the

X/TARE button again. The meter shows the last previously stored 4-digit number (-1999 through

9999) with flashing 4

th

digit.

4. Press

the

S/NT/GRS button to change the digits.

5. Press

the

X/TARE button to scroll to the next digit.

6. Press

the

MENU button to store your selection.

(Output 1) displays.

This starting analog signal corresponds to your Read 1 display.

7. Press

the

X/TARE button. Selected output displays.

If you select

for voltage, the maximum signal you may select is 10.00 for a 0-10 Vdc

signal output. If you select

for current, the maximum signal you may select is 20.00.

8. Press

the

S/NT/GRS button to enter the Output 1 signal selection. If you continue to press the S/NT/GRS

button, the flashing digit’s value continues to change.

9. Press

the

X/TARE button to scroll horizontally to the next digit.

10. Press the MENU button to store your selection.

(Read 2) displays.

This is your second point of display reading.

11. Press the

X/TARE button. The meter shows the last previously stored 4-digit number (-1999 from 9999) with

flashing 4

th

digit.

12. Press the

S/NT/GRS button to change the value of the flashing digit. If you continue to press the S/NT/GRS

button, the flashing digit’s value continues to change.

13. Press the

X/TARE button to scroll horizontally to the next digit.

14. Press the MENU button to store your selection. The meter shows

(Output 2).

This analog signal should correspond to your Read 2 display.

15. Press the

X/TARE button. The meter shows the selected output.

If you select

for voltage, the maximum signal you may select is 10.00 for a 0-10 V signal output.

If you select

for current, the maximum signal you may select is 20.00 for a 0-20 or 4-20 mA DC

signal output.

16. Press the

S/NT/GRS button to change the value of the flashing digit. If you continue to press the S/NT/GRS

button, the flashing digit’s value continues to change.

17. Press the

X/TARE button to scroll horizontally to the next digit.

18. Press the MENU button to store your selection. The meter momentarily shows

, followed by

(Lock Out configuration).

WARNING: If the meter shows flashing values on any time, the value has overflowed. Press the
S/NT/GRS button to start new values.

4.11.1 Examples for Output Scale and Offset

Example: You want to spend 4-20 mA output for 0 to 100.0. The meter has 0.1 degrees resolution. Complete the
following steps:
1. Press

the

MENU button until the meter shows

.

2. Press

the

X/TARE button. The meter shows

(Read 1).

3. Press

the

X/TARE button to show the existing value.

4. Change the value of Read 1 to 000.0 by pressing the

S/NT/GRS and X/TARE buttons.

5. Press

the

MENU button to store your selection. The meter shows

(Output 1).

6. Press

the

X/TARE button to show the existing value.

7. Change the value of Output 1 to 04.00 by pressing the

S/NT/GRS and X/TARE buttons.

8. Press

the

MENU button to store your selection. The meter shows

(Read 2).