DFI INFINITY CS (V- M3598/N/0906)

29

CF 70

ACCURACY AT 25

o

C

Max Error Strain:

±0.03% of reading, ± 1 count

Span

Tempco:

50

ppm/°C

Step

Response:

1

sec

Warm Up to Rated Accuracy:

30 min

Excitation Voltage:

AC power units

DC power units

24

V

@

25

mA,

Refer to Table 8-1

12 V @ 50 mA,

Color chart for DC Output

10

V

@

120

mA, Excitation

5

V

@

60

mA

Load Regulation:

1.1%

Line Regulation:

0.02% per Vac

ALARM OUTPUTS (if applicable)

2 Form “C” on/off relays. Configurable for latched and unlatched by software.

Max current: 5 AMPS, resistive load

Max voltage: 250 Vac or 30 Vdc

ANALOG OUTPUT (if applicable)

Signal Type:

Current or voltage

Signal Level:

Current: 10 V max compliance at 20 mA output

Voltage: 20 mA max for 0-10 V output

Function:

May be assigned to a display range or proportional control output with setpoint #1
when used as a control output.

Linearity: 0.2%
Step Response Time:

2 – 3 seconds to 99% of the final value

ISOLATED ANALOG OUTPUT (TB5, if applicable)

Same as non-isolated analog output except isolated.

Signal Type:

Current or voltage

Signal Level:

Current: 10 V max compliance at 20 mA output

Voltage: 20 mA max for 0-10 V output

Function:

May be assigned to a display range or proportional control output with Setpoint #1
when used as a control output.

Linearity: 0.2%
Step Response Time:

2 – 3 seconds to 99% of the final value

Isolation:

130 Vrms working voltage, 1000 V/60sec Dielectric test

Only one analog output is available on each unit and it must be factory installed.

INPUT POWER INFORMATION

AC units

115/230 V~(AC) ±10%, 50/60 Hz

9.5

W

max,

power

consumption (Non-Isolated Analog Out)

11.0 W max, power consumption (Isolated Analog Out)

DC Units

10-32 Vdc or 26-56 Vdc, 8 W
Do not use a combination of dc power and internal excitation or Isolated Analog
Out, unless using dc power of 20-32 Vdc.

Refer

to

Table

8-1

below

External Fuse Required:





IEC 127-2/111
Power Fuse
115V

125 mA @ 250 (T)

230V

63 mA @ 260 (T)

UL 248-14 (Listed Fuse)
Power Fuse
115V

175 mA @ 250 V Slow-Blow

230V

80 mA @ 250 Slow-Blow