Cooper Instruments & Systems DFI Infinity CS Economy Digital Indicator User Manual

Page 26

Advertising
background image

DFI INFINITY CS (V- M3598/N/0906)

23

CF 70

9. Press

the

X/TARE button to show the existing value.

10. Change the value of Read 2 to 100.0 by pressing the

S/NT/GRS and X/TARE buttons.

11. Press the MENU button to store your selection. The meter shows

(Output 2).

12. Press the

X/TARE button to show the existing value.

13. Change the value of Output 2 to 20.0 by pressing the

S/NT/GRS and X/TARE buttons.

14. Press the MENU button to store your selection. The meter shows

(Lock Out Configuration).

4.12 Using Lock Out Configuration

Refer to Table 6-1 for a summary list of menu configuration.

Use Lock Out Configuration

for the following:

• To enable or disable setpoint changes

• To enable or disable the RESET button in the Run Mode

• To enable or disable displaying meter firmware version

4.12.1 Enabling or Disabling the RESET button in the Run Mode

1. Press

the

MENU button until the meter shows

(after

.)

2. Press

the

X/TARE button. The meter shows one of the following:

= To enable the RESET button in the Run mode (Default)

= To disable the RESET button in the Run mode

3. Press

the

S/NT/GRS button to toggle between available choices.

4. Press

the

MENU button to store the changes. The meter shows

if the new value is different otherwise

the meter shows

and returns to the Run mode.

4.12.2 Enabling or Disabling the SETPOINT Changes

1. Press

the

MENU button until the meter shows

(after

)

2. Press

the

X/TARE button twice. The meter shows one of the following:

= To enable the setpoint changes (Default)

= To disable the setpoint changes

3. Press

the

S/NT/GRS button to toggle between available choices.

4. Press

the

MENU button to store the changes. The meter shows

if the new value is different otherwise

the meter shows

and returns to the Run Mode.

4.12.3 SETPOINT Display Function: Firmware version or Setpoint value

1. Press

the

MENU button until the meter shows

(after

)

2. Press

the

X/TARE button three times. The meter shows one of the following:

= SETPTS button will display setpoint values.

= SETPTS button will display the meter’s firmware version.

3. Press

the

S/NT/GRS button to toggle between the choices above.

4. Press

the

MENU button to store the changes. The meter shows

if the new value is different otherwise

the meter shows

and returns to the Run Mode.

If your meter does not have the relay option, setpoint menu items above will not be available and SETPTS
button will always display the meter’s firmware version. These units will have

(overload) or

memory indicated by Alarm 1 & 2 LED displays. LEDs can be reset by pressing MENU then RESET
button or by Power OFF then ON.

4.13 Using Display Brightness Configuration

4.13.1 Changing Brightness Level

Changing Display Brightness is not active unless your meter is a Version “B”.
1. Press

the

MENU button the meter shows

(after

).

2. Press

the

X/TARE button from

. The meter shows one of the following: