Maintenance

|

15

Use the cleaning tape shipped with the HI-6105 kit for cleaning the end

face of the connector. Softly press the connector face on the exposed

cleaning tape and move in a figure eight motion. After three

figure eights inspect with a microscope and repeat if necessary.

The mating connectors such as bulkhead feedthrough, transmit, and

receive ports can be cleaned with canned dry air. Fiber optic

connectors should be cleaned every time they are mated and

unmated.

H

ANDLING

I

NSTRUCTIONS

Never touch the end face of the connector.

Connectors not in use should be covered over the ferrule by a plastic dust cap or

boot. These boots will protect the polished ferrule end from damage but not

contamination. It is important to note that inside the dust cap there is a sticky

residue that is a by-product of making the dust cap. This residue will remain on

the ferrule end after the cap is removed. Always clean fiber connectors and

mating connectors prior to installation.

Upgrade Policies

System upgrades are made periodically to enhance functionality. Contact

ETS-Lindgren Customer Service or see

www.ets-lindgren.com

to determine the

upgrade status of your measurement system.

Service Procedures

For the steps to return a system or system component to ETS-Lindgren for

service, see the Product Information Bulletin included with your shipment.