Read battery voltages, Read peak, Reset peak – ETS-Lindgren HI-6100 Field Monitor User Manual

Page 51

Advertising
background image

Remote Operation

|

51

—Read Battery Voltages

Returns the current battery percentages of probes in use.

Syntax: RBAT?

Parameters: None

To check the current battery percent of the probes in use: RBAT?<LF>

Response: 62%,100%,,<CR><LF>

Channels 3 and 4 are not on or not ready.

—Read Peak

Writes the peak probe field values to a remote port.

Syntax: RDPK,x

Parameters: (x)

1 = Reads channel 1 max

5 = Reads channel 1 min

2 = Reads channel 2 max

6 = Reads channel 2 min

3 = Reads channel 3 max

7 = Reads channel 3 min

4 = Reads channel 4 max

8 = Reads channel 4 min

Output format: Ch1 Max xx.xx,Xxx.xx,Yxx.xx,Zxx.xx<CR><LF>

—Reset Peak

Resets peak probe field values.

Syntax: RSPK

Parameters: None