Read battery voltages, Read peak, Reset peak – ETS-Lindgren HI-6100 Field Monitor User Manual
Page 51
Advertising

Remote Operation
|
51
—Read Battery Voltages
Returns the current battery percentages of probes in use.
Syntax: RBAT?
Parameters: None
To check the current battery percent of the probes in use: RBAT?<LF>
Response: 62%,100%,,<CR><LF>
Channels 3 and 4 are not on or not ready.
—Read Peak
Writes the peak probe field values to a remote port.
Syntax: RDPK,x
Parameters: (x)
1 = Reads channel 1 max
5 = Reads channel 1 min
2 = Reads channel 2 max
6 = Reads channel 2 min
3 = Reads channel 3 max
7 = Reads channel 3 min
4 = Reads channel 4 max
8 = Reads channel 4 min
Output format: Ch1 Max xx.xx,Xxx.xx,Yxx.xx,Zxx.xx<CR><LF>
—Reset Peak
Resets peak probe field values.
Syntax: RSPK
Parameters: None