7.0 Error Handling and Troubleshooting

Error Handling

The user interface display contains two error list boxes. These error list boxes

may be displayed from the DISPLAY menu by selecting LOWER DISPLAY. If

any errors have occurred since the system was powered on, each box will

display up to the last six error messages.

Troubleshooting

Following are some common troubleshooting tips for the HI-6100 Field Monitor. If

you have a problem not listed or if the corrective action fails, contact

ETS-Lindgren Customer Service.

Indication

Possible Cause / Corrective Action

Probe fails to

communicate with

HI-6100

Fiber optic cable reversed: Confirm cable is

connected properly

Dirty fiber optic cable ends: Clean to remove dirt

or contaminants

Bad cable or faulty connector: Replace cable

Dead battery: Charge battery

HI-6100 fails to

power up

Power cord unplugged or faulty: Confirm cable

is connected properly. To replace cord, contact

Customer Service.