0 error handling and troubleshooting, Error handling, Troubleshooting – ETS-Lindgren HI-6100 Field Monitor User Manual
Page 57

Error Handling and Troubleshooting
|
57
7.0 Error Handling and Troubleshooting
Error Handling
The user interface display contains two error list boxes. These error list boxes
may be displayed from the DISPLAY menu by selecting LOWER DISPLAY. If
any errors have occurred since the system was powered on, each box will
display up to the last six error messages.
Troubleshooting
Following are some common troubleshooting tips for the HI-6100 Field Monitor. If
you have a problem not listed or if the corrective action fails, contact
ETS-Lindgren Customer Service.
Indication
Possible Cause / Corrective Action
Probe fails to
communicate with
HI-6100
Fiber optic cable reversed: Confirm cable is
connected properly
Dirty fiber optic cable ends: Clean to remove dirt
or contaminants
Bad cable or faulty connector: Replace cable
Dead battery: Charge battery
HI-6100 fails to
power up
Power cord unplugged or faulty: Confirm cable
is connected properly. To replace cord, contact
Customer Service.