Ieee-488 (gpib) communications, Rs-232 communications, Rs-232 settings – ETS-Lindgren HI-6100 Field Monitor User Manual

Page 48: Figure 34, Rs-232 communication settings table

Advertising
background image

48

|

Remote Operation

IEEE-488 (GPIB) Communications

For General Purpose Interface Bus (GPIB) communications, the End or

Identify (EOI) control line may also be used for command termination. Terminate

the command with <LF>, EOI, or both when sending commands to the receiver

via the GPIB. No characters are permitted after <LF> or EOI; the system

interprets characters following <LF> or EOI as the start of another command.

When an error condition is present at the receiver, the service request (SRQ)

signal is asserted. The operator can then perform a serial poll operation. The

receiver error code (in binary) is contained in the lower bits of the serial poll

status byte. If the receiver is addressed as a listener and the GPIB remote line is

asserted, the receiver will switch to remote mode.

RS-232 Communications

Commands are processed in a manner very similar to that of the IEEE interface.

The command structure is identical, except that an EOI line does not exist;

therefore, <LF> must terminate all commands. When a valid command is

received, it is processed and the result is immediately transmitted back over the

RS-232 interface. The output format is identical to that of the IEEE format except,

again, an EOI line does not exist.

RS-232 Settings

Word Length

8 bits

Stop Bits

1

Baud Rate

1200

– 115,200 (menu selectable)

Parity

None

Figure 34:

RS-232 Communication Settings Table