1.0 Introduction

The ETS-Lindgren HI-6100 Field Monitor is a broadband electric and magnetic

field probe monitor designed for use in radio frequency

interference/electromagnetic compatibility (RFI/EMC) test system applications. It

accepts inputs from up to four isotropic field probes and analyzes and displays

that information on a user-configurable LCD display. It also provides two digital

interfaces: IEEE-488 and RS-232.

Figure 1: HI-6100 Field Monitor

The HI-6100 is compatible with ETS-Lindgren electric (E) field or magnetic (H)

field probes or laser probes. It has a probe-dependent frequency response of

10 kHz to 40 GHz and a probe-dependent sensitivity of 0.5 V/m to 3000 V/m,

15.0 mA/m to 30 A/m.

The HI-6100 incorporates a safety interlock mechanism that turns off the laser if

it does not receive data from the probe within a specified time frame. The safety

mechanism is intended to prevent injury from the laser if the HI-6100 issues a

command to turn on the laser while the fiber optic cables are disconnected,

improperly connected, cut, or damaged.

The HI-6100 in native HI-6100 mode is the preferred method for use with the

laser-powered field probes as well as the HI-60XX Series battery-powered field

probes.

The HI-6100 in FM5004-compatible mode is primarily used with the HI-44XX

Series probes for backward portability.