Message Codes

Leica DISTO™ plus 1.0.0 GB

de

GB

fr

it

es

nl

ja

*

Maximum deviation occurs under unfavorable condi-
tions, such as in bright sunlight or when measuring to
inadequately reflecting surfaces. For distances over
30m-without using a target plate- themaximum devia-
tion may increase by ±0.1mm/m to maximum ±
10mm. For more information, refer to www.disto.com.

** As of 100m the measuring units are displayed in cm.
*** Reduced when using the BLUETOOTH

®

function

Message Codes

Error message

appears beside the message number on the
display.

Please call your local dealership and specify the
message code. Should the message continue to
appear after several starts, then the instrument
is defective.

Care

Look after the optical surfaces with the same care that you
would apply to spectacles, cameras and field glasses.
The most current information can be found on our
homepage:
www.disto.com
We publish the addresses of specialized software
mannufacturers . However, we are not responsible for the
correctness of their specifications or for the performance
of their software.
For software developers, the Leica DISTO™ plus Online-
Interface description is available.

All illustrations, descriptions and technical specifica-
tions are subject to change without prior notice.

Message

Code

Cause

Remedy

203

Wrong sequence with

Pythagoras meas.

Carry out measurement in

correct sequence

204

Calculation error

Repeat procedure

240

Transmission error

Repeat procedure

252

Temperature too high,

above 50°C (measuring)

Cool down instrument

253

Temperature too low,

below -10°C (measuring)

Warm up instrument

255

Receiver signal too weak,

measurement time too

long, distance <200 mm

Use target plate

measurement time >10

sec.

256

Received signal too

powerful

Use target plate (correct

side)

257

Wrong measurement,

ambient brightness too

high

Use target plate

260

Laser beam interrupted

Repeat measurement

All other messages

Call service "System"

Message

Code

Cause

Remedy