Measurement data should not be collected when the %Torque reading is out of range. The

out of range condition can be resolved by either changing the speed (reduce speed when

reading is out of range: high) or changing the spindle (increase the spindle size when the

reading is out of range: low).

TIP: When comparing data, the test method is critical. Be sure that you know the proper

spindle and speed required for the test method. If readings are out of range, this

condition should be reported as the test result.

II.8 Printing

The DV2T Viscometer can communicate to a Dymo Label Writer 450 Turbo label printer.

This printer can be purchased from Brookfield (Part No. GV-1046). The communication

to the printer is by USB (cable provided with the printer). When the printer is connected

to the DV2T, the printer icon will be visible in the status bar.

The DV2T Viscometer can configure the print out for several formats of paper/labels.

These various paper/label stocks are available from Brookfield.

1.

GV-1049-10 SHIPPING LABEL 2.31 X 4IN. 300 P/ROLL (PKG of 10)

2.

GV-1048-10 ADDRESS LABEL 1.13 X 3.5IN. 350 P/ROLL (PKG of 10)

3.

GV-1047-10 CONTINUOUS PAPER 2.25IN. WD. X 300FT. LG (PKG of 10)