Brookfield DV2T Viscometer User Manual

Page 9

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 9

Manual No. M13-167-B0614

Electrical Certifications:

Conforms to CE Standards:

BSEN 61326:

Electrical equipment for measurement, control and laboratory use

- EMC requirements

BSEN 61010-1: Safety requirements for electrical equipment, for measurement,

control and laboratory use

Notice to customers:

This symbol indicates that this product is to be recycled at an appropriate collection center.
Users within the European Union:

Please contact your dealer or the local authorities in charge of waste management on how

to dispose of this product properly. All Brookfield offices and our network of representatives

and dealers can be found on our website: www.brookfieldengineering.com

Users outside of the European Union:

Please dispose of this product according to your local laws.

I.4 Installation

Note:

“IQ, OQ, PQ”, an abbreviated guideline document for installation, operation and

performance validation for your DV2T digital viscometer can be downloaded

from our website www.brookfieldengineering.com. A more detailed IQ,OQ,PQ

procedure is available for purchase from Brookfield or your authorized dealer.

1) Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix I).

2) Put the viscometer on the stand.

3) Connect the RTD probe to the socket on the rear panel of the DV2T.

4) The Viscometer must be leveled. The level is adjusted using the two leveling screws

on the base. Adjust the leveling screws so that the bubble level on the front of the

DV2T is centered within the circle.

Note

: Check level periodically during use.

5) Remove the shipping cap which secures the coupling nut on the Viscometer to the

pivot cup. For Cone/Plate Models, hold the Sample Cup and swing the tension

bar away from the bottom of the cup. Lower the cup and remove the foam insert.

(Save for future shipments.)

6) Optional

: Install the screen protector per the instructions on the package (and also

shown in Appendix K). Failure to properly install the screen protector may result

in touch screen malfunction. Additional installation help can also be found on

our YouTube channel: www.youtube.com/user/BrookfieldEng

The AC input voltage and frequency must be within the appropriate range as

shown on the nameplate of the viscometer (see Section I.2).