IV. SETTINGS

The Settings menu provides access to the many controls and features of the DV2T Viscometer.

This menu can be accessed through the Settings Navigation Icon

which is often present

in the Title Bar. Figure IV-1 shows the Settings Menu which is divided into: Device Setup,

User Settings, Global Settings, and Admin Functions. Global Settings include items that affect

the complete range of features within the DV2T. Admin Functions include items related to

administrator level controls.

Figure IV-1

IV.1 Device Setup

The Device Setup menu includes settings related to the mechanical systems and connections

of the DV2T. Six settings are available including:

Temperature - Create off sets to be used with specific temperature probes.

Printer Setup

- Define the print format to be used with the Dymo 450 Printer.

AutoZero

- Force the DV2T to perform an AutoZero as is done at start up.

Oscillation Check - Evaluate the performance of the DV2T lower bearings.

Technical Support Info - Information that may be requested by Brookfield Technical

Support staff to assist with troubleshooting.

About

- Present basic information about the DV2T as is done at start up.

Temperature:

The Temperature setting allows the user to create up to 10 temperature

offset values for connected temperature probes. Brookfield offers several

types of temperature probes for use with the DV2T including: DVP-94Y

immersion probe (supplied with the DV2T), CPA-44PYZ embedded probe

(optional with Cone/Plate DV2T Viscometers), and SC4-13RP embedded

probe (optional with Small Sample Adapter). Any of these probes can be

calibrated locally against a standard reference thermometer to determine

an offset (how far from the actual temperature does the probe read).

This offset can be entered into the DV2T Viscometer and identified with

a name defined by the user.