Brookfield DV2T Viscometer User Manual

Page 37

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 37

Manual No. M13-167-B0614

Materials that exhibit thixotropy will show a steadily decreasing measured viscosity over

time. Materials that exhibit pseudoplasticity will show a changing viscosity as the spindle

speed changes.

TIP: When averaging data for a thixotropic material, begin the Averaging Duration after

the period of most significant change in measured viscosity. This will reduce the

variability in the averaged value.

TIP: When averaging data for a pseudoplastic material, do not average together (Test

Averaging) data collected at different speeds (or shear rates).

Live Averaging of data occurs during actual testing of a sample. Data can be collected

as an average of readings over a specific time interval; each data point saved in the file is

an averaged value. This averaging is defined in the Data Collection section of Configure

Viscosity Test.

Single Point Averaging

requires a definition of Averaging Duration, the amount of time

for which readings will be averaged. This time parameter will be applied at the end of

the End Condition. In this case a single data point will be collected which represents

the average of all data measured during the specified time period (Averaging Duration).
Multi Point Averaging

requires a definition of 1) Data Interval, the frequency of data

collection and 2) Averaging Duration, the amount of time for which readings will be

averaged. These two parameters will work in conjunction to generate multiple data

points each of which represent the average of all data measured during the average

duration within the specified Data Interval.

Post Test Averaging of data occurs after the test is complete, through the Results screen.

Averages and standard deviation values can be generated for data collected in a single

step or across several steps of a test generated through PG Flash software. If the test

used to collect data utilized Live Averaging (described above), then the Post Test Averaging

will produce an average value of averaged data points.

Post Test Averaging can be specified in the Report section of Configure Viscosity Test (see

Section III.7). When configured as part of the test, the average values will be displayed

immediately upon test conclusion. If not specified in Report, Post Test Averaging can

be selected from the Down Arrow in the Results screen (see Section III.9).
Post Test Averaging offers two options: Step Averaging and Test Averaging (see Figure

III-13).

Figure III-13

Advertising