Page 90

Manual No. M13-167-B0614

Viscometer Will Not Return to Zero

Viscometer is not level

• Check with spindle out of the sample.

• Adjust the laboratory stand.

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

Remove the spindle and turn the motor OFF; select display to % torque mode.

Gently push up on the viscometer coupling.

Turn the coupling until the digital display reads 10-15 on the % display.

Gently let go of the coupling.

Watch the digital display; you should see a “run” of numbers next to the %; the

“run” of numbers

should ultimately stop at 0.0 (+/- 0.1).

If the digital display does not return to ZERO, the unit most likely is in need of

service.

• Perform calibration check (see Appendix F).

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair (see Appendix O).

* This procedure does not apply to instruments with ball bearing suspension (see Section

I.3).

Display Reading Will Not Stabilize

Special characteristic of sample fluid. There is no problem with the viscometer.

• Refer to Appendix C.

Check for erratic spindle rotation.

• Verify power supply

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

✓ Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

Temperature fluctuation in sample fluid.

• Use temperature bath for control.