IV.4.5 Default Path

The Path is the location where data or tests are stored. The DV2T will allow for storage

to the Internal Memory or to an attached USB Flash drive. The path will also include any

file structure that has been created.
The Default Path identifies what location will be used as the initial location when Results are

saved. The Default Path defined in Reports - Configure Viscosity Test (saved test or

unsaved test) will take priority over the Default Path.
The DV2T can create file structure based on User ID and Test Name. The options provided

for the Default Path specify what format the data path should utilize. For example: set

the Default Path to Operator -> Test Name. The User is logged in as DAVID and the test

to run is INK 5. The data path will be defaulted as shown in Figure IV-10.

Figure IV-10

TIP: The Default Path can be overwritten at the time that data is saved.

IV.4.6 Settings Reset

Settings Reset will return the DV2T settings (Device, User and Global) to the factory default.

Internal memory will not be affected. User accounts will not be affected.

IV.4.7 Device Reset

DATA LOSS POSSIBLE!

Device Reset will return the DV2T to the factory default condition. All data, tests and Audit

Trails will be deleted from internal memory. All User accounts will be deleted.

TIP: Move data files and test files to a USB USB Flash Drive prior to selecting Device

Reset.