Brookfield Engineering Labs., Inc.

Page 59

Manual No. M13-167-B0614

A.2 Setup

1. Be sure that the Viscometer is securely

mounted to the Laboratory Stand, leveled and

zeroed with no cone or cup attached and 0%

torque is displayed.

2. Figure A-2

shows a typical water bath setup.

Connect the sample cup inlet/outlet ports to the

water bath inlet and outlet and set the bath to

the desired test temperature. Allow sufficient

time for the bath to reach the test temperature.

3. The Viscometer has been supplied with a

special cone spindle(s) which contains the

Electronic Gap Setting feature. The “CPE” or

“CPA” part number designation on the cone

spindle verifies the Electronic Gap Setting
feature.

Note

:

The “CPE” or “CPA” cone or

cup cannot be used with earlier

DV2T cone/plate Viscometers

(below S/N50969) which do not

have the electronic gap setting

feature.

4. With the motor off, thread the cone spindle

by using the spindle wrench to secure the

viscometer coupling nut (see Figure A-3);

gently push up on the coupling nut and hold

this securely with the wrench. Thread the cone

spindle by hand.

Note

:

Left Hand Threads.

5. Attach the cup, taking care not to hit the cone

with the cup (Figure A-4), by positioning the

cup against the mic ring and swinging the

tension bar under the cup. The tension bar

should have plastic tubing in place.

Bath/Circulator

Bath

Inlet

Bath

Outlet

Sample

Cup

Cup

Outlet

Cup
Inlet

Figure A-3

Spindle

Wrench

Cone Spindle

These surfaces

must be clean!

Coupling Nut

Micrometer

Adjustment

Ring

Do Not hit the

CONE with the CUP!

Tension Bar

Figure A-4

Figure A-2