Maintenance and Backup


NucleoView® NC-200™ Software Maintenance

A two-channel image taken with NucleoView® NC-200™ is approximately 12 MB in size. Therefore, for
example taking 50 images each work day will over a month accumulate approximately 18 GB.
The user must hence backup results and free up disc space in good time to ensure trouble free operation.
See Image Files for further details about file types and locations.

The NucleoView® NC-200™ constantly records all operations in event log files. These event log files should
be handled in the same manner as the image files with respect to backing up and freeing up disc space. The
event log files are always stored in the current results folder. One file is created for each month, and files
are named according to the rule <year><month>.log like 201005.log indicating a log file from May 2010.

The NucleoView® NC-200™ software including standard protocols is typically delivered on a USB stick, and
protocols and licenses are distributed as zip-files via e-mail, USB stick or other media.
The user must keep appropriate backup of these zip-files in order to be able to make a new installation in
case re-installation of the software is required.


NucleoCounter® NC-200™ Instrument Maintenance

ChemoMetec A/S provides specific Installation Qualification (IQ), Operation Qualification (OQ) and
Performance Qualification (PQ) protocols and associated test kits, which may be used to verify proper
operation of the NucleoView® and NucleoCounter® NC-200™ system.

The NucleoCounter® NC-200™ instrument can be cleaned in case it has been contaminated with sample
fluids, dust or other spillage. Use dry cleaning swabs for cleaning, and only if this is not enough, use 96-
100% pure ethanol. Under no circumstances should the inside of the cassette fixture tried to be cleaned.
The outer surfaces of the NucleoCounter® NC-200™ instrument may also be cleaned with 96-100% pure
ethanol if needed.

It is good practice to put the large lid down, whenever the instrument is not being used, to prevent dust
build up on the internal surfaces. Also avoid splashing sample fluids into the interior of the instruments, as
cleaning of these parts should be done only by ChemoMetec A/S.