Trouble shooting – ChemoMetec NC-200 User Manual
Page 46
 
991-0204 Rev. 1.1
September 2012
46
Trouble Shooting 
 
 
Installation 
The NucleoView® NC-200™ 
software or the NucleoCounter® 
NC-200™ instrument is not 
installed correctly. 
 
The user must be logged on with administrator rights during 
installation. 
 
When connecting the NucleoCounter® NC-200™ instrument to a 
previously unused USB port, the NucleoView® NC-200™ software may 
need to reinstall the camera drivers and will prompt the user to do so. 
This operation may need administrator rights, so either use another 
USB port where the camera has already been installed, or log on with 
administrator rights and complete the camera installation as 
described in Installation of Instrument. 
 
 
NucleoView® NC-200™ Issues 
(Software) 
Protocol file is invalid 
 
 
 
Connection Error 
 
 
 
The license file is corrupted, missing or does not include a license for 
the selected protocol. The appropriate license must be installed as 
described in the Licenses section. 
 
The NucleoCounter® NC-200™ instrument icon in the upper left 
corner of the main window stays red after NucleoView® NC-200™ has 
been launched, i.e. the NucleoView® NC-200™ software does not 
connect to the instrument. This may be caused by an error in the USB 
connection. To inspect and solve this, open the Control Panel and 
launch System. Select the Hardware tab and click Device Manger. In 
case of a USB driver error there will be a USB item marked with a 
yellow dot containing an exclamation mark. Select properties for this 
item and select the driver tab. Select Update Driver and let the system 
select a driver automatically.