Cassette Issues
No cassette is found 
 
 
 
Non-valid cassette 
 
 
 
Flow detection error 
 
This may be because either no cassette is in the instrument or the 
cassette is placed incorrectly. In the latter case, remove the cassette 
and re-insert it. 
 
Typically the user has placed an already used cassette in the 
instrument. Simply replace with an unused cassette which has been 
correctly loaded. 
 
The flow has failed to reach the flow detection point within given time 
limits. A common cause is that too small a volume of sample solution 
has been loaded into the cassette. Simply place a cassette which has 
been correctly loaded. Another typical cause is that the user has 
placed an already used cassette in the instrument. 
NucleoCounter® NC-200™ Issues 
(Instrument) 
Abnormal sounds
 
 
Typically the piston motor has been blocked by a cassette that has not 
been correctly inserted. Stop the run by removing the power supply. 
Reconnect the power supply and remove the cassette and re-insert it 
again before starting a new run. 
 