3 percent mode (%) – A&D Weighing FX-3000i User Manual

Page 18

Advertising
background image

16

5-3 Percent Mode (%)

This is the mode to display the weight value in percentage compared to a 100% reference mass and

is used for target weighing or checking the sample variable.

Selecting the percent mode

1 Press the MODE key to select % (percent mode).

Storing the 100% reference mass

2 Press the SAMPLE key to enter the 100% reference

mass storing mode.

Even in the storing mode, pressing the MODE key will

switch to the next mode.

3 Place a container on the weighing pan, if necessary.

Press the RE-ZERO key to cancel the weight (tare). The
balance displays 100 0

%

.

4 Place the sample to be set as the 100% reference mass

on the pan or in the container.

5 Press the PRINT key to store the reference mass. The

balance displays 100.00

%

. (The decimal point position

depends on the reference value. The reference mass
stored, even if the AC adapter is removed, is maintained
in non-volatile memory.)

Note
If the balance judges that the mass of the sample is too
light to be used as a reference, it displays
lo . Do not use
the sample.

6 Remove the sample.

Reading the percentage

7 Place a sample to be compared to the reference mass on

the pan. The displayed percentage is based on the 100%
reference mass.

While the stabilization indicator is on, pressing the PRINT
key will output the weight value, using the RS-232C serial
interface.

Note
Peripheral equipment, that is sold separately , such
as a printer or a personal computer is required.

Advertising