5-5 Statistical Calculation Mode (Example of Use)

Here, as an example of use of the statistical calculation mode, mixing of the multiple formulae such
as medicine is described. The mixing process is recorded using the balance and the printer.
In the example, the FX-3000i and the AD-8121B (dump print mode) are connected using the RS-232C
serial interface.

5-5-1 Getting Started

Changing the function table

Changes :

●To enable the statistical calculation mode

●To enable “Zero after output”

Enabling the statistical calculation mode

1 Enter the function table menu.
Press and hold the SAMPLE key until ba5fnc of

the function table is displayed, then release the key.


2 Select the application function.

Press the SAMPLE key several times to display

1

ap fnc . Then, press the PRINT key to display

1

°apf 0 .


3 Change the application function parameter to “2”.
Press the RE-ZERO key to display apf 2 . Press the

PRINT key to confirm the change. After end ,

1

ba5fnc

1

is displayed.

Enabling “Zero after output”

4 Select “Zero after output”.
Press the SAMPLE key several times to display

1

dout .

Then, press the PRINT key to display

°prt 0 , and press

the SAMPLE key several times to display

1

°ar-d 0 .


5 Enable “Zero after output”.
Press the RE-ZERO key to display ar-d 1 . Then,

press the PRINT key to confirm the change. After

1

end , 5if is displayed.

Returning to the weighing mode

6 Press the CAL key to return to the weighing mode.