7-3 Correcting the Internal Mass Value (Only for FZ-i series)

The FZ-i series balance can correct the internal mass value by basing on an external weight.
The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed.
(The FX-i series balance does not have the internal mass value correction function.)
Select an appropriate weight as listed below:

Model

Usable calibration weight

Adjustable range

FZ-120

i

100 g, 50 g

FZ-200

i

200 g, 100 g, 50 g

FZ-300

i

300

g,

200 g, 100 g, 50 g

-0.050 g to +0.050 g

FZ-1200

i

1000 g, 500 g

FZ-2000

i

2000 g, 1000 g, 500 g

FZ-3000

i

3000

g,

2000 g, 1000 g, 500 g

-0.50 g to +0.50 g

The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.

Operation

1 With factory setting, the balance can not correct the internal

mass value. Refer to “8. FUNCTION SWITCH AND
INITIALIZATION” and set the function and the internal mass
value correction function.


2 With the weighing mode, press and hold the SAMPLE key to

display ba5fnc .


3 Press the SAMPLE key several times until Cal5et is

displayed.

If Cal5et is not displayed, perform the step 1.


4 Press the PRINT key to display Cal 0 .

Refer to “7-4 Calibration Using an External Weight”, to perform
the calibration.


5 After calibration, the balance displays ready

?

.

After placing the small break breeze on the balance, press the

PRINT key.

6 The balance corrects the internal mass value automatically

after displaying Cal5et .

7 After correcting the internal mass value, the balance displays

C

Calin , and performs the calibration automatically using the

corrected internal mass value.

8 The

balance

displays ba5fnc after displaying end .


9 Press the CAL key to return the weighing mode. Confirm that the

internal mass value has been corrected properly by placing the
mass used to performed the correction, on the weighing pan.

If the internal mass value has not been corrected properly,

repeat the procedure, starting at step 2 and adjust the
correction value.

Press and

hold

Place the

mass

Place the

small break

breeze

Adjusting

the internal

mass value
Calibrating

the internal

mass value