28

5 Confirm that there is nothing on the pan and

press the PRINT key. The balance measures the
zero point. Do not allow vibration or drafts to affect
the balance.

The balance displays the calibration weight value.

6 Place the displayed calibration weight on the pan

and press the PRINT key. The balance measures
the calibration weight. Do not allow vibration or
drafts to affect the balance.

7 The balance displays end . Remove the weight

from the pan.

8 When the GLP output is set, the balance displays

1

glp and outputs “Calibration Report”.

9 The balance will automatically return to the

weighing mode.

10 Place the calibration weight on the pan and confirm

that the value displayed is within

±2 digits of the

specified value. If it is not within the range, check
the ambient conditions such as breeze and
vibration, also check the weighing pan. Then,
repeat steps 1 to 9.