7-2 Calibration Using the Internal mass

(One-Touch

Calibration, only for

FZ-

i

series

)

This function calibrates the balance using the internal mass. The only operation required is to press

the CAL key

Operation

1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the

weighing pan.

2 After carefully installing the small breeze break,

press the CAL key.

3 The balance displays Calin and performs calibration using the internal mass. Do not allow

vibration or drafts to affect the balance.

4 The balance displays end after calibration. If the “GLP output (info)” parameter of the

function table is set to “1” or “2”, the balance displays glp and outputs “Calibration Report”
using the RS-232C interface or stores the data in memory. For details on the calibration report
format, refer to “10-2 GLP Report”.

5 The balance will automatically return to the weighing mode after calibration.

About the internal mass

The mass of the internal mass may change due to corrosion or other damage caused by the
operating environment, or due to aging. Check the internal mass periodically. Correct the internal
mass value as necessary. For details, refer to “7-3 Correcting the internal mass value”.
To maintain the weighing accuracy, perform the calibration using an external weight periodically,
as described below.